GETINGE 46-SERIES 46-4

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06,07 report with the FDA on 2013-08-22 for GETINGE 46-SERIES 46-4 manufactured by Getinge Disinfection Ab.

Event Text Entries

[15198329] Smoke was observed coming from a getinge model 46-4 washer disinfector during use. The customer shut off the unit. No injuries or damage to the facility have been reported.
Patient Sequence No: 1, Text Type: D, B5


[15603493] The device eval is currently in-progress. A voluntary device correction of getinge model46-4 water disinfectors was reported to us fda on 07/26/2013. This correction was initiated to address a software function that can be utilized during servicing of the unit. This is a password protected function but if left in the enabled state, a user could inadvertently activate the heating element without water in the chamber which could lead to an overheat condition. The device correction actions include inspecting that this function is not enabled and to installed a heating element with integral thermostat to address the delayed overheat protection. The incident subject of this mdr is still being investigated to determine its cause and to evaluate if any additional corrective actions are required as a result of this incident. No flame was reported in this instance though electrical wiring and components were damaged in the washer disinfector. No damage to the facility or injury to individuals were reported. A f/u mdr will be submitted at the conclusion of this incident investigation.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9616031-2013-00003
MDR Report Key3370958
Report Source06,07
Date Received2013-08-22
Date of Report2013-07-23
Date of Event2013-07-22
Device Manufacturer Date2006-08-01
Date Added to Maude2013-11-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactANN WHEELER, US AGENT
Manufacturer Street1777 E. HENRIETTA RD.
Manufacturer CityROCHESTER NY 14623
Manufacturer CountryUS
Manufacturer Postal14623
Manufacturer Phone5852725036
Manufacturer G1GETINGE DISINFECTION AB
Manufacturer StreetLJUNGADALSGATAN 11
Manufacturer CityVAXJO 35115
Manufacturer CountrySW
Manufacturer Postal Code351 15
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameGETINGE 46-SERIES
Generic NameWASHER DISINFECTOR
Product CodeMDZ
Date Received2013-08-22
Model Number46-4
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerGETINGE DISINFECTION AB
Manufacturer AddressLJUNGADALSGATAN 11 VAXJO 35115 SW 351 15


Patients

Patient NumberTreatmentOutcomeDate
10 2013-08-22

© 2026 FDA.report