GENTLE TOUCH SYS 100MM WITH LOOP OSTOMY ROD 020934

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06,07 report with the FDA on 2013-09-09 for GENTLE TOUCH SYS 100MM WITH LOOP OSTOMY ROD 020934 manufactured by Convatec, A Division Of E.r. Squibb & Sons L.l.c..

Event Text Entries

[3747313] Operating room manager with unity health system reports that this morning a surgeon opened a gentle touch loop ostomy system in the operating room, and just prior to inserting the rod, the swivel tip snapped off. The rod had no contact with the patient. Both pieces are to be returned along with the gentle touch package.
Patient Sequence No: 1, Text Type: D, B5


[11171455] The loop rod sample was broken along the long axis. (b)(4) was raised. The retain sample was examined. Components were present, assembled properly and consistent with product specification requirements. There was no evidence of any mfg defect. Components used were approved for use. The loop rod was examined and the swivel end checked. The swivel end was rotated 10 cycles and each time clicked to correct location with no evidence of any stress to the swivel or loop rod. The batch record review found no objective evidence of discrepancies related to the complaint issue. Case was identified during a recent fda facility inspection. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1049092-2013-00008
MDR Report Key3371524
Report Source05,06,07
Date Received2013-09-09
Date of Report2011-06-10
Date of Event2011-06-10
Date Mfgr Received2011-06-10
Date Added to Maude2013-12-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMARY SZARO
Manufacturer Street200 HEADQUARTERS PARK DRIVE.
Manufacturer CitySKILLMAN NJ 08558
Manufacturer CountryUS
Manufacturer Postal08558
Manufacturer Phone9089042450
Manufacturer G1CONVATEC, A DIVISION OF E.R. SQUIBB & SONS L.L.C.
Manufacturer Street211 AMERICAN AVE.
Manufacturer CityGREENSBORO NC 27409
Manufacturer CountryUS
Manufacturer Postal Code27409
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameGENTLE TOUCH SYS 100MM WITH LOOP OSTOMY ROD
Generic NameROD, COLOSTOMY
Product CodeEZP
Date Received2013-09-09
Returned To Mfg2011-07-28
Model Number020934
Catalog Number020934
Lot NumberOH02364
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Sequence No1
Device Event Key0
ManufacturerCONVATEC, A DIVISION OF E.R. SQUIBB & SONS L.L.C.
Manufacturer Address211 AMERICAN AVE. GREENSBORO NC 27409 US 27409


Patients

Patient NumberTreatmentOutcomeDate
10 2013-09-09

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.