RELIANCE 1227 CART WASHER

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2013-09-26 for RELIANCE 1227 CART WASHER manufactured by Steris Canada Corporation.

Event Text Entries

[18853541] The user facility reported a gust of hot air escaped the washer when the door was opened which initiated the sprinkler system. No injuries were reported. No procedural delays or cancellations occurred.
Patient Sequence No: 1, Text Type: D, B5


[19293748] A steris service technician inspected the unit and found the low speed pressure switch not working properly. The technician replaced the low pressure switch, ran a test cycle and confirmed the unit operational. A hospital employee used the service security code to override the locking mechanism to open the washer door, which caused the reported steam to infiltrate the room. The 1227 cart washer operator manual does not instruct use of the service security code; the service security code is to be used by steris technicians only. The hospital staff has been has been counseled that they are not, under any circumstances, to use the service security code to override a cycle or alarm that the unit may be experiencing.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9680353-2013-00129
MDR Report Key3371559
Report Source06
Date Received2013-09-26
Date of Report2013-09-26
Date of Event2013-08-28
Date Mfgr Received2013-08-28
Date Added to Maude2014-01-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS KATHRYN CADORETTE
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927231
Manufacturer G1STERIS CANADA CORPORATION
Manufacturer Street490, ARMAND-PARIS
Manufacturer CityQUEBEC, QUEBEC GIC8A3
Manufacturer CountryCA
Manufacturer Postal CodeGIC 8A3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRELIANCE 1227 CART WASHER
Generic NameWASHER
Product CodeFLH
Date Received2013-09-26
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTERIS CANADA CORPORATION
Manufacturer Address490 ARMAND-PARIS QUEBEC, QUEBEC GIC8A3 CA GIC 8A3


Patients

Patient NumberTreatmentOutcomeDate
10 2013-09-26