VITROS CHEMISTRY PRODUCTS GENT REAGENT 6801711

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2013-09-26 for VITROS CHEMISTRY PRODUCTS GENT REAGENT 6801711 manufactured by Ortho-clinical Diagnostics.

Event Text Entries

[3747349] The customer obtained multiple lower than expected vitros gentamicin (gent) quality control results (qc fluid results 5. 16, 4. 66, 5. 07 versus expected 7. 51 ug/ml) on a vitros 5,1 fs chemistry system. Biased results of the magnitude and direction observed may lead to inappropriate physician action if patient samples were affected. No patient samples were known to have been affected or reported from the laboratory. There was no allegation of patient harm as a result of this event. This report corresponds to ortho clinical diagnostics inc. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[11228156] The investigation determined that lower than expected quality control results were obtained for vitros gent on a vitros 5,1 fs chemistry system. The investigation could not determine a definitive cause. Precision testing performed using an alternate gent reagent lot suggests that an instrument issue is not a likely contributing factor. An unknown quality issue with the affected gent reagent lot cannot be ruled out as a contributing factor. Expected performance was obtained using an alternate vitros gent reagent lot.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1319808-2013-00043
MDR Report Key3371789
Report Source05
Date Received2013-09-26
Date of Report2013-09-26
Date of Event2013-08-13
Date Mfgr Received2013-08-28
Device Manufacturer Date2013-02-12
Date Added to Maude2014-01-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. JOSEPH FALVO
Manufacturer Street100 INDIGO CREEK DRIVE
Manufacturer CityROCHESTER NY 14626
Manufacturer CountryUS
Manufacturer Postal14626
Manufacturer Phone5854533000
Manufacturer G1ORTHO-CLINICAL DIAGNOSTICS
Manufacturer Street1000 LEE ROAD
Manufacturer CityROCHESTER NY 14606
Manufacturer CountryUS
Manufacturer Postal Code14606
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVITROS CHEMISTRY PRODUCTS GENT REAGENT
Generic NameIN-VITRO DIAGNOSTIC
Product CodeLCD
Date Received2013-09-26
Catalog Number6801711
Lot Number1512-08-2760
Device Expiration Date2014-11-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerORTHO-CLINICAL DIAGNOSTICS
Manufacturer Address100 INDIGO CREEK DRIVE ROCHESTER NY 14626 US 14626


Patients

Patient NumberTreatmentOutcomeDate
10 2013-09-26

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.