NUSTAR DENTAL CHAIR DC1700

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 08 report with the FDA on 2013-09-03 for NUSTAR DENTAL CHAIR DC1700 manufactured by Marus.

Event Text Entries

[3933884] A dentist had positioned a pt for treatment in a lowered reclined position when the lift assembly broke on the marus dc1700 dental chair resulting in the seat assembly to fall backwards towards the floor with the pt in the chair. The pt valid out of the back of the chair and onto the floor. There were no injuries reported.
Patient Sequence No: 1, Text Type: D, B5


[11263244] Upon eval it was identified that the two welded support block on the front truss assembly had failed as a result of improper welding from our supplier as it was not welded to manufacturers specifications. Marus contacted the supplier of the front truss assembly to inform them of this issue. The supplier informed marus that they will define areas of work in process (wip) for status identification of the welding process. They will also update fqc document to include a conformity check of weldments to ensure front truss assemblies are manufactured to marus specifications. All front truss assemblies the supplier had in inventory were to marus specifications. Marus also confirmed all front truss assemblies in house from our supplier were to manufacturers specifications.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1017522-2013-00015
MDR Report Key3376662
Report Source08
Date Received2013-09-03
Date of Report2013-08-07
Date of Event2013-08-07
Date Mfgr Received2013-08-07
Device Manufacturer Date2013-06-01
Date Added to Maude2013-12-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactFRANK RAY, MGR
Manufacturer Street11727 FRUEHAUF DR.
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal28273
Manufacturer Phone7045877227
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameNUSTAR DENTAL CHAIR
Generic NameDENTAL CHAIR
Product CodeKLC
Date Received2013-09-03
Returned To Mfg2013-08-09
Model NumberDC1700
Catalog NumberDC1700
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMARUS
Manufacturer Address11727 FRUEHAUF DR. CHARLOTTE NC 28273 US 28273


Patients

Patient NumberTreatmentOutcomeDate
10 2013-09-03