SMART MONITOR 2 PS 1014557

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2013-09-04 for SMART MONITOR 2 PS 1014557 manufactured by Philips Respironics - Chmv.

Event Text Entries

[3767581] Children's medical ventures (chmv) received a report from a durable medical equipment (dme) supplier stating that a smart monitor infant apnea monitor had failed the self-test. No resulting effect on a pt or potential user has been reported. The device was reportedly not in use at the time of the reported event. However, it is not known if the device alarmed appropriately during the self-test process. The dme has been contacted for additional info and stated that they were not aware of a complaint allegation associated with the device. Additionally, the dme stated that the device's location was not currently known but that it was not currently with a pt. No additional info is available at this time. The alleged failure is being reported due to the fact that it is unk whether or not the unit alarmed appropriately for the reported event.
Patient Sequence No: 1, Text Type: D, B5


[11193550] Pma/510(k): k061256. (b)(4). The complaint issue alleged by the customer was not able to be confirmed because the device has not been returned to the mfr for eval. The purpose of the smartmonitor 2 functional self-test is to check that all the features of the unit are functioning properly. A functional self-test should be performed at least once a week or according to the instructions given by the health care professional. You should also perform the test: after a lead wire is changed; after the pt cable is changed. The user manual states the following user/owner responsibilities. The respironics monitor and accessories are designed to work as described in the operator's manual. The user(s) of the equipment should not use parts that have failed, exhibit excessive wear, are contaminated, or otherwise ineffective. The monitor and its accessories should not be modified. The smartmonitor2 ps and smartmonitor2 psl are intended for use in the continuous monitoring of respiration, heart rate, and spo2 levels (smartmonitor2 ps only) of infant, pediatric, and adult pts. The device detects and alarms for periods of high or low heart rate, high or low breath rate, and high or low saturation (smartmonitor2 ps only). When used as an infant monitor, it is intended for use in a home hospital environment. For infants only, it monitors and alarms for central apneas. When used as a pediatric or adult monitor, it is intended for use in a hospital environment. Pt alarm limits are set by the health care professional before the smartmonitor2 ps is delivered to the pt and are typically set up the prescribed settings which include a delay before recording apneas and a delay before annunciating an alarm for an apnea condition. Based on a complete review of the complaint allegation, chmv has no reason to believe the device in question did not perform to specification; effectively notifying the caregiver of a potential event. It is determined that no further investigation into the alleged complaint issue is appropriate at this time. However, if the device is returned for investigation and it is concluded that a malfunction outside of the design specifications took place, a f/u additional info report will be filed.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3007056120-2013-00014
MDR Report Key3382845
Report Source05
Date Received2013-09-04
Date of Report2013-08-05
Date Mfgr Received2013-08-05
Device Manufacturer Date2011-05-01
Date Added to Maude2013-12-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCHRIS FERGUSON
Manufacturer Street191 WYNGATE DR
Manufacturer CityMONROEVILLE PA 15146
Manufacturer CountryUS
Manufacturer Postal15146
Single Use3
Previous Use Code3
Removal Correction NumberNO
Event Type3
Type of Report3

Device Details

Brand NameSMART MONITOR 2 PS
Product CodeNPF
Date Received2013-09-04
Model Number1014557
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerPHILIPS RESPIRONICS - CHMV
Manufacturer Address191 WYNGATE DR MONROEVILLE PA 15146 US 15146


Patients

Patient NumberTreatmentOutcomeDate
10 2013-09-04

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.