HUDSON SHERIDAN/HVT TRACHEAL TUBE, 8.5 5-13017

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,08 report with the FDA on 2013-09-10 for HUDSON SHERIDAN/HVT TRACHEAL TUBE, 8.5 5-13017 manufactured by Teleflex Medical.

Event Text Entries

[18866539] The event is reported as: the customer alleges having difficulty in inflating the cuff of the endotracheal tube. No pt injury/involvement.
Patient Sequence No: 1, Text Type: D, B5


[19064116] From the picture it was observed that "difficult to inflate cuff. " no other defects were observed. A dimensional and functional inspection of the product involved in the complaint could not be conducted since the product was not returned. The dhr (device history record) review could not be conducted since the lot number was not provided. A document assessment (fmea) was conducted and no changes were required. From the picture it was observed "difficult to inflate cuff", the complaint can not be confirmed and a conclusive root cause can not be determined since the sample was not available. Teleflex will continue to monitor and trend relating complaints.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3003898360-2013-00404
MDR Report Key3390575
Report Source01,08
Date Received2013-09-10
Date of Report2013-08-30
Date of Event2013-07-29
Date Mfgr Received2013-08-30
Date Added to Maude2013-12-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMARGIE BURTON, RN
Manufacturer StreetPO BOX 12600
Manufacturer CityDURHAM NC 27709
Manufacturer CountryUS
Manufacturer Postal27709
Manufacturer Phone9194334965
Manufacturer StreetPROLONGACION MISION EUSEBIO KINO # 1316,RANCHO EL DESCANSO
Manufacturer CityTECATE, B.C. 21478
Manufacturer CountryMX
Manufacturer Postal Code21478
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHUDSON SHERIDAN/HVT TRACHEAL TUBE, 8.5
Generic NameENDOTRACHEAL TUBE
Product CodeLNZ
Date Received2013-09-10
Catalog Number5-13017
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerTELEFLEX MEDICAL
Manufacturer AddressTECATE, B.C. MX


Patients

Patient NumberTreatmentOutcomeDate
10 2013-09-10