MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05 report with the FDA on 2013-10-07 for ARCHITECT C8000 SYSTEM 01G06-01 manufactured by Abbott Manufacturing Inc.
[3753347]
The customer reports a falsely elevated potassium assay result generated for one patient sample on the architect c8000 analyzer. An initial result of 7. 1 mmol/l was generated and not reported from the lab. The sample retested at 4. 7, 4. 7 and 4. 8 mmol/l. There is no impact to patient management reported.
Patient Sequence No: 1, Text Type: D, B5
[11263185]
An evaluation is in process. A follow-up report will be submitted when the evaluation is complete. (b)(4).
Patient Sequence No: 1, Text Type: N, H10
Report Number | 1628664-2013-00277 |
MDR Report Key | 3391078 |
Report Source | 01,05 |
Date Received | 2013-10-07 |
Date of Report | 2013-09-25 |
Date Mfgr Received | 2013-09-25 |
Date Added to Maude | 2013-12-24 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | NOEMI ROMERO-KONDOS, RN BSN |
Manufacturer Street | 100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3 |
Manufacturer City | ABBOTT PARK IL 600643537 |
Manufacturer Country | US |
Manufacturer Postal | 600643537 |
Manufacturer Phone | 847937-512 |
Manufacturer G1 | ABBOTT LABORATORIES |
Manufacturer Street | 100 ABBOTT PARK ROAD |
Manufacturer City | ABBOTT PARK IL 60064350 |
Manufacturer Country | US |
Manufacturer Postal Code | 60064 3500 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | ARCHITECT C8000 SYSTEM |
Product Code | MZV |
Date Received | 2013-10-07 |
Catalog Number | 01G06-01 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Eval'ed by Mfgr | N |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ABBOTT MANUFACTURING INC |
Manufacturer Address | 1921 HURD DRIVE IRVING TX 75038 US 75038 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|