METRX EX0513120

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2013-10-07 for METRX EX0513120 manufactured by Medtronic Sofamor Danek Usa, Inc.

Event Text Entries

[3752907] It was reported that the tip of the curette broke off in the patient disc space during a discectomy. The disc space collapsed. The tip of the currette was left in the disc space.
Patient Sequence No: 1, Text Type: D, B5


[11224478] (b)(6). (b)(4). Location : hospital. The device has been returned to the manufacturer for evaluation. Macroscopic examination confirms ~5mm of probe tip broken off, the broken off portion of the tip is missing and not returned for analysis. Microscopic examination of the fracture revealed a quasi-brittle fracture surface with no indications of torsion or fatigue. Plastic deformation of tip suggests the direction of fracture, consistent with bend stress overload. Tip diameter and material hardness inspected and found to be conforming to print specification. The above observations are consistent with bend stress overload.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1030489-2013-04091
MDR Report Key3391092
Report Source05
Date Received2013-10-07
Date of Report2013-09-11
Date of Event2013-09-11
Date Mfgr Received2013-09-11
Device Manufacturer Date2013-08-22
Date Added to Maude2013-10-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactHUZEFA MAMOOLA
Manufacturer Street1800 PYRAMID PLACE
Manufacturer CityMEMPHIS TN 38132
Manufacturer CountryUS
Manufacturer Postal38132
Manufacturer Phone9013963133
Manufacturer G1MEDTRONIC SOFAMOR DANEK
Manufacturer Street1800 PYRAMID PLACE
Manufacturer CityMEMPHIS TN 38132
Manufacturer CountryUS
Manufacturer Postal Code38132
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMETRX
Generic NameCURETTE
Product CodeHTF
Date Received2013-10-07
Returned To Mfg2013-10-02
Model NumberNA
Catalog NumberEX0513120
Lot NumberTI13H045
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC SOFAMOR DANEK USA, INC
Manufacturer Address4340 SWINEA RD MEMPHIS TN 38118 US 38118


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2013-10-07

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.