CUSA EXCEL 36KHZ STRAIGHT HANDPIECE C2602

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05 report with the FDA on 2013-09-18 for CUSA EXCEL 36KHZ STRAIGHT HANDPIECE C2602 manufactured by Integra Lifesciences (ireland) Limited.

Event Text Entries

[3800323] This is the third report of three involving a malfunction that occurred during a procedure on the same pt at the same facility (cross reference with mfr 3006697299-2013-00065 (b)(4) and mfr 3006697299-2013-00070 (b)(4)). The theatre reported of a 36khz handpiece getting very hot and the small of burning after only a few seconds of using. This blackened the end of the tip. They replaced all consumables; however, handpiece and tip still very very hot. Diathermy not working properly either. The machine was also making a high pitched screeching noise. The unit was in contact with a pt. The procedure was delayed an hour but, the pt did not incur an injury as a result of this event. The sales rep visited the account and performed a full check with and without the valleylab force fx nosecone attached to diathermy machine. The console, handpiece and tip were 100% serviceable. The sales rep also took advise from the engineer to ensure she completed all necessary checks. After 25 minutes of constant use the machine still continued to work perfectly.
Patient Sequence No: 1, Text Type: D, B5


[11282450] To date, the device involved in the reported incident has not been received for eval. An investigation has been initiated based upon the reported info.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3006697299-2013-00071
MDR Report Key3394785
Report Source01,05
Date Received2013-09-18
Date of Report2013-09-18
Date of Event2013-08-16
Date Mfgr Received2013-08-20
Date Added to Maude2013-12-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactLINDA SERENTINO
Manufacturer Street315 ENTERPRISE DRIVE
Manufacturer CountryUS
Manufacturer Phone6099365560
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCUSA EXCEL 36KHZ STRAIGHT HANDPIECE
Generic NameULTRASONIC SURGICAL PRODUCTS
Product CodeLBK
Date Received2013-09-18
Catalog NumberC2602
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerINTEGRA LIFESCIENCES (IRELAND) LIMITED
Manufacturer AddressSRAGH, TULLAMORE, CO. OFFALY EI


Patients

Patient NumberTreatmentOutcomeDate
10 2013-09-18