TERUMO CDI 100 MONITOR

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2013-09-13 for TERUMO CDI 100 MONITOR manufactured by Terumo Cardiovascular Systems Corp..

Event Text Entries

[19393313] This complaint was confirmed by lab inspection. The unit's battery was cracked and leaking internal fluid leading to the corrosion of the battery compartment. According to the analysis, this prevented the unit from starting up. The unit would start when associated power supply was used, with all monitor functions working to specification. Service and replacement parts are no longer available. No add'l action will be taken at this time.
Patient Sequence No: 1, Text Type: N, H10


[19482152] It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the monitor would shut down unit. The device was not changed out, as the customer ran the case without hematocrit saturation module (h/sat). The surgical procedure was completed successfully, and there were no delays, no blood loss or no adverse consequences to the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1828100-2013-00877
MDR Report Key3396422
Report Source05,06
Date Received2013-09-13
Date of Report2013-09-11
Date of Event2013-04-08
Date Mfgr Received2013-09-11
Date Added to Maude2013-12-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJAN WINDER
Manufacturer Street6200 JACKSON RD.
Manufacturer CityANN ARBOR MI 48103
Manufacturer CountryUS
Manufacturer Postal48103
Manufacturer Phone7346634145
Manufacturer G1TERUMO CARDIOVASCULAR SYSTEMS CORP.
Manufacturer Street6200 JACKSON RD.
Manufacturer CityANN ARBOR MI 48103
Manufacturer CountryUS
Manufacturer Postal Code48103
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTERUMO CDI 100 MONITOR
Product CodeDTY
Date Received2013-09-13
Returned To Mfg2013-05-13
Model Number100
Catalog Number100
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTERUMO CARDIOVASCULAR SYSTEMS CORP.
Manufacturer Address6200 JACKSON RD. ANN ARBOR MI 48103 US 48103


Patients

Patient NumberTreatmentOutcomeDate
10 2013-09-13

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.