[245834] Rptr was trying to secure the button into the specimen holder, which required some force to snap-in to the slot. During this maneuver, their fingers slipped from the button and hit the knife blade, resulting in a cut on right middle finger. Compared to other cryostats, this particular model requires snap-in holder of the button and has a large portion of the blade being exposed to give a potential for cutting injury.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW1022337
MDR Report Key340960
Date Received2001-07-05
Date of Report2001-07-05
Date of Event2001-04-03
Date Added to Maude2001-07-11
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA0
Report to FDA0
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameCLINICUT 60 CRYOSTAT
Generic NameMICROTOME CRYOSTAT
Product CodeIDN
Date Received2001-07-05
Model NumberCLINICUT 60 CRYOSTAT
Catalog Number*
Lot Number*
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key330287
ManufacturerHACKER INSTRUMENTS & INDUSTRIES
Manufacturer Address17 SHERWOOD LANE PO BOX 10033 FAIRFIELD NJ 07004 US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2001-07-05