MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2001-07-05 for CLINICUT 60 CRYOSTAT * manufactured by Hacker Instruments & Industries.
[245834]
Rptr was trying to secure the button into the specimen holder, which required some force to snap-in to the slot. During this maneuver, their fingers slipped from the button and hit the knife blade, resulting in a cut on right middle finger. Compared to other cryostats, this particular model requires snap-in holder of the button and has a large portion of the blade being exposed to give a potential for cutting injury.
Patient Sequence No: 1, Text Type: D, B5
Report Number | MW1022337 |
MDR Report Key | 340960 |
Date Received | 2001-07-05 |
Date of Report | 2001-07-05 |
Date of Event | 2001-04-03 |
Date Added to Maude | 2001-07-11 |
Event Key | 0 |
Report Source Code | Voluntary report |
Manufacturer Link | N |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 0 |
Health Professional | 3 |
Initial Report to FDA | 0 |
Report to FDA | 0 |
Event Location | 3 |
Single Use | 0 |
Previous Use Code | 0 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | CLINICUT 60 CRYOSTAT |
Generic Name | MICROTOME CRYOSTAT |
Product Code | IDN |
Date Received | 2001-07-05 |
Model Number | CLINICUT 60 CRYOSTAT |
Catalog Number | * |
Lot Number | * |
ID Number | * |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Implant Flag | N |
Date Removed | A |
Device Sequence No | 1 |
Device Event Key | 330287 |
Manufacturer | HACKER INSTRUMENTS & INDUSTRIES |
Manufacturer Address | 17 SHERWOOD LANE PO BOX 10033 FAIRFIELD NJ 07004 US |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Other | 2001-07-05 |