HOOK 165MM 2.0MM 90DEG BLUNT [MXI/XOM] MCO805

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06,07 report with the FDA on 2013-10-21 for HOOK 165MM 2.0MM 90DEG BLUNT [MXI/XOM] MCO805 manufactured by Xomed Microfrance Mfg.

Event Text Entries

[3911083] It was reported the distal extremity of the hook broke inside the ear canal during surgery. No reported patient impact.
Patient Sequence No: 1, Text Type: D, B5


[11121429] This device is used for therapeutic purposes. The batch number (lot number) of the instrument is not known but the design indicates that the instrument is at least 10 years old. (b)(4): the tip of the hook is broken. The broken fragment is missing. The observation of the breakage zone has not highlighted any manufacturing or material defects and indicates that the break has been sudden. Considering the age and the absence of manufacturing or material defect, the most probable cause is a weakening of the instrument during it years of use that has led to the breakage observed with a shock / constraint during use.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9680837-2013-00487
MDR Report Key3420610
Report Source01,05,06,07
Date Received2013-10-21
Date of Report2013-10-02
Date of Event2013-09-24
Date Mfgr Received2013-10-02
Date Added to Maude2014-01-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCHRISTY CAIN
Manufacturer Street6743 SOUTHPOINT DRIVE NORTH
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal32216
Manufacturer Phone9043328353
Manufacturer G1MEDTRONIC XOMED, INC.
Manufacturer Street6743 SOUTHPOINT DRIVE NORTH
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal Code32216
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHOOK 165MM 2.0MM 90DEG BLUNT [MXI/XOM]
Generic NameHOOK, MICROSURGICAL EAR
Product CodeJYL
Date Received2013-10-21
Returned To Mfg2013-10-02
Model NumberMCO805
Catalog NumberMCO805
Lot NumberSEE H10
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerXOMED MICROFRANCE MFG
Manufacturer AddressSAINT-AUBIN-LE-MONIA BOURBON-L?ARCHAMBAULT 3160 FR 3160


Patients

Patient NumberTreatmentOutcomeDate
10 2013-10-21