MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00,01,06 report with the FDA on 2013-10-14 for PORT A CATH II INTRASPINAL IMPLANTABLE ACCESS SYSTEM 21-1500 manufactured by Smiths Medical Inc..
[3895886]
From clinical trial study organizer: it was reported that the device was implanted in patient for administration of clinical trial drug on (b)(6) 2011. According to report, after an unknown amount of time in use, it was difficult to obtain cerebrospinal fluid (csf) return from the device; flush of device and drug delivery were completed without issue. The device was surgically explanted on (b)(6) 2013: the removed portion of catheter was found to have a broken outlet tube which demonstrated a leak during flush of the device. No permanent adverse effects to patient reported.
Patient Sequence No: 1, Text Type: D, B5
[11339402]
The device is currently being evaluated; the mfr will file a follow up reported detailing the results of the evaluation once it is completed.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 2183502-2013-00602 |
MDR Report Key | 3429103 |
Report Source | 00,01,06 |
Date Received | 2013-10-14 |
Date of Report | 2013-10-14 |
Date of Event | 2013-09-11 |
Date Facility Aware | 2013-09-11 |
Report Date | 2013-10-14 |
Date Reported to FDA | 2013-10-14 |
Date Mfgr Received | 2013-09-16 |
Device Manufacturer Date | 2010-04-01 |
Date Added to Maude | 2013-10-28 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | PETE HIRTE |
Manufacturer Street | 1265 GREY FOX RD. |
Manufacturer City | ST. PAUL MN 55112 |
Manufacturer Country | US |
Manufacturer Postal | 55112 |
Manufacturer Phone | 6516287384 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | PORT A CATH II INTRASPINAL IMPLANTABLE ACCESS SYSTEM |
Generic Name | CATHETER, PERCUTANEOUS, LONG TERM, INTRASPINAL |
Product Code | LNY |
Date Received | 2013-10-14 |
Returned To Mfg | 2013-10-07 |
Model Number | 21-1500 |
Catalog Number | 21-1500 |
Lot Number | 1768497 |
ID Number | NA |
Device Expiration Date | 2015-04-01 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | 42 YR |
Device Eval'ed by Mfgr | N |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | SMITHS MEDICAL INC. |
Manufacturer Address | ST. PAUL MN US |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Required No Informationntervention | 2013-10-14 |