SUR-FIT NATURA 2 PC-POST OP DRAINABLE KIT 416930

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,04,05,07 report with the FDA on 2013-10-21 for SUR-FIT NATURA 2 PC-POST OP DRAINABLE KIT 416930 manufactured by Convatec Dominican Republic, Inc..

Event Text Entries

[3901260] It is reported that during installation of the loop rod in the operating room, the nozzle of the loop rod fell off. As a result, the articulation of the loop rod withdrew during the passage into the colon of the muscular wall, supported by a lever mechanism. It is further reported that the small removable part is back in the abdomen.
Patient Sequence No: 1, Text Type: D, B5


[11445744] Based on the info provided this event is deemed serious injury. The following clinical consequences for the pt is reported as follows: very inflammatory and edematous colon and that the operation was extended for more than 15 minutes. Further info reports that the period between the date on which the sales representative became aware of the complaint and the date of notification to the (b)(4) resulting in pending an appointment with the surgeon to have the necessary info for this report, and the completed form by the complainant. No additional pt/event details have been provided to date. A return sample for evaluation is not expected. Should additional info becomes available a follow-up report will be submitted. Reported to the fda on (b)(4) 2013. Note: the actual date of event is unk, so the date used was the date convatec became aware.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9618003-2013-00042
MDR Report Key3429916
Report Source01,04,05,07
Date Received2013-10-21
Date of Report2013-09-10
Date of Event2013-09-10
Date Mfgr Received2013-09-10
Date Added to Maude2013-10-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMARY SZARO, ASSOC. DIRECTOR
Manufacturer Street200 HEADQUARTERS PARK DRIVE
Manufacturer CitySKILLMAN NJ 08558
Manufacturer CountryUS
Manufacturer Postal08558
Manufacturer Phone9089042450
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSUR-FIT NATURA 2 PC-POST OP DRAINABLE KIT
Generic NameROD, COLOSTOMY
Product CodeEZP
Date Received2013-10-21
Model Number416930
Catalog Number416930
Lot NumberCNO
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCONVATEC DOMINICAN REPUBLIC, INC.
Manufacturer AddressCARRETERA SANCHEZ, KM. 18.5 PARQUE INDUSTRIAL ITABO, S.A. HAINA, SAN CRISTOBAL 33102 DR 33102


Patients

Patient NumberTreatmentOutcomeDate
10 2013-10-21

© 2026 FDA.report