BLUE SPRINGS WP-280R

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 1996-06-03 for BLUE SPRINGS WP-280R manufactured by Blue Springs Corp..

Event Text Entries

[27130] Pt expired following myocardial infarction and cardiac arrest approx 90 mins after a hemodialysis treatment. The pt's dialysis was terminated early due to a suspected pyrogen reaction. Subsequent investigation indicates that the reverse osmosis system (ro) used to treat the water for reprocessing dialyzers had a ruptured membrane. Reverse osmosis pump had failed, and pieces of the broken pump were found in the ruptured reverse osmosis membrane. It is unknown as to whether there is any association between the suspected pyrogen reaction and the pt death. Seven other pts also experienced suspected pyrogen reactions during the same morning: six of whom recovered and left the facility that morning; the seventh pt was hospitalized overnight for observation and was discharged the next day having recovered with no intervention.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number34327
MDR Report Key34327
Date Received1996-06-03
Date of Report1996-06-03
Date of Event1996-05-17
Date Facility Aware1996-05-17
Report Date1996-06-03
Date Reported to FDA1996-06-03
Date Reported to Mfgr1996-06-03
Date Added to Maude1996-07-24
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameBLUE SPRINGS
Generic NameREVERSE OSMOSIS SYSTEM
Product CodeJRS
Date Received1996-06-03
Model NumberWP-280R
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age11 YR
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key35686
ManufacturerBLUE SPRINGS CORP.
Manufacturer AddressENCINO CA 91436 US


Patients

Patient NumberTreatmentOutcomeDate
101. Death 1996-06-03

© 2026 FDA.report