CONNECTPOINT

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2013-11-01 for CONNECTPOINT manufactured by Steris Corporation - Montgomery.

Event Text Entries

[17156183] The user facility reported that an anesthesiologist bent down to the floor and in the process of standing up hit his head on the harmony connectpoint. The anesthesiologist was sent to the er. The user facility declined to provide additional event information. No procedural delays/cancellations were reported.
Patient Sequence No: 1, Text Type: D, B5


[17366976] A steris service technician arrived onsite and inspected the connectpoint. The technician adjusted the connectpoint to several heights in a 45 degree range; the connectpoint stayed in each position. The technician tightened the brakes on the system and found that more pressure was needed to adjust arm up and down. When properly adjusted, the arm did not move down. The system was turned back to service and no further issues have been reported. Steris is filing this report due to the absence of information regarding the reported injury.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1043572-2013-00110
MDR Report Key3442959
Report Source06
Date Received2013-11-01
Date of Report2013-11-01
Date of Event2013-10-02
Date Mfgr Received2013-10-02
Date Added to Maude2013-11-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS KATHRYN CADORETTE
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927231
Manufacturer G1STERIS CORPORATION - MONTGOMERY
Manufacturer Street2720 GUNTER PARK DRIVE EAST
Manufacturer CityMONTGOMERY AL 36109
Manufacturer CountryUS
Manufacturer Postal Code36109
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCONNECTPOINT
Generic NameINTEGRATION PACKAGE CUSTOM
Product CodeLMD
Date Received2013-11-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTERIS CORPORATION - MONTGOMERY
Manufacturer Address2720 GUNTER PARK DRIVE EAST MONTGOMERY AL 36109 US 36109


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2013-11-01

© 2026 FDA.report