MEPITEL ONE 289670-00

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00 report with the FDA on 2013-10-30 for MEPITEL ONE 289670-00 manufactured by Molnlycke Health Care.

Event Text Entries

[3783083] (b)(6) boy had friction burn on the back of his hand that had been grafted during surgery, mepitel one was applied to the graft and then a secondary dressing on top. The boy was brought back to theatre 48 hours later, saline was poured over the dressing, which made the safetac tacky, the dressing was then removed, which also removed the skin graft.
Patient Sequence No: 1, Text Type: D, B5


[11375877] Based on the available information, this event is deemed a serious injury. The use of mepitel one for fixating grafts is within the intended use. There is a causal link to the product, although removal of the dressing already after 48 hours is deemed far too early. This is supported by the precautions located in the product instructions for use. "when mepitel one is used for fixation of skin grafts, the dressing should not be changed before the fifth day post application. " no additional patient/ event information has been provided to date. A return sample for evaluation is not expected. Should additional information become available, a follow up report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004763499-2013-00018
MDR Report Key3452561
Report Source00
Date Received2013-10-30
Date of Report2013-10-29
Date of Event2013-10-22
Date Mfgr Received2013-10-22
Date Added to Maude2013-11-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactANGELA BUNN, DIRECTOR
Manufacturer Street5550 PEACHTREE PARKWAY SUITE 500
Manufacturer CityNORCROSS GA 30092
Manufacturer CountryUS
Manufacturer Postal30092
Manufacturer Phone6782507930
Manufacturer G1MOLNLYCKE HEALTHCARE OY
Manufacturer StreetSAIMAANUKATU 6 FI-50101
Manufacturer CityMIKKELI 10
Manufacturer CountryFI
Manufacturer Postal Code10
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMEPITEL ONE
Generic NameMEPITEL ONE
Product CodeMGP
Date Received2013-10-30
Model NumberNA
Catalog Number289670-00
Lot Number13288780
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMOLNLYCKE HEALTH CARE
Manufacturer AddressNORCROSS GA US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2013-10-30

© 2026 FDA.report