[16552954] Siemens u. S. Was notified on september 2, 2013 that during a service call on (b)(4) 2013, siemens' customer service engineer (cse) rotated the gantry by using the hand control. However, when the button to stop rotation was pressed, the motion switch was released and the same command given on the touch screen, the gantry rotation did not stop. Reportedly, the cse pressed the yellow emergency button at the hand control but the gantry rotation still did not stop. Then, the cse turned off the f1 fuse to dis-energize the system; however, it had already "crashed. " there is no report of mistreatment or injury to a patient. This reported issue occurred in (b)(6).
Patient Sequence No: 1, Text Type: D, B5


[16709020] Siemens u. S. Became aware of the reported event on september 2, 2013. This mdr is being mailed on september 24, 2013. Siemens' investigation into the reported event is ongoing. A supplemental report will be submitted to the fda upon completion of the investigation. Customer address: (b)(6). Brand name: simview nt, p/n 5496992 and simview nt with idi, p/n 5497016 are both affected products for the reported problem.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2910081-2013-07739
MDR Report Key3456084
Report Source01,07
Date Received2013-09-25
Date of Report2013-07-24
Date of Event2013-07-24
Date Mfgr Received2013-07-24
Date Added to Maude2014-01-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMARLYNNE GALLOWAY
Manufacturer Street51 VALLEY STREAM PKWY, MS D02
Manufacturer CityMALVERN PA 19355
Manufacturer CountryUS
Manufacturer Postal19355
Manufacturer Phone6102195361
Manufacturer G1SIEMENS MEDICAL SOLUTIONS USA, INC.
Manufacturer Street757A ARNOLD DR
Manufacturer CityMARTINEZ CA 94553
Manufacturer CountryUS
Manufacturer Postal Code94553
Single Use0
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSIMVIEW NT WITH IDI
Generic NameSYSTEM, SIMULATION, RADIATION THERAPY
Product CodeKPQ
Date Received2013-09-25
Model Number05497016
Catalog NumberNA
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS MEDICAL SOLUTIONS USA, INC., RADIATION ONCOLOGY
Manufacturer Address757A ARNOLD DR MARTINEZ CA 94553 US 94553


Patients

Patient NumberTreatmentOutcomeDate
10 2013-09-25