VARIANTNBS WITH GDM 3.1 SOFTWARE 250-3000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,07 report with the FDA on 2013-10-15 for VARIANTNBS WITH GDM 3.1 SOFTWARE 250-3000 manufactured by Bio-rad Laboratories.

Event Text Entries

[3731439] On (b)(6) 2013, a customer located in (b)(6) was using the variant nbs sickle cell program and noticed that a sample ((b)(6)) appeared twice on the summary report. Within the same run, a sample result ((b)(6)) showed a flat line with no peaks on the chromatogram. On (b)(6) 2013, the operator repeated the entire run to further investigate the results that appeared twice on the summary report as well as the flat line chromatogram. The summary report from the second run contained no repeated samples and all chromatograms showed peaks (no flat lines). The incident was reported to bio-rad labs.
Patient Sequence No: 1, Text Type: D, B5


[11375491] For preliminary investigation bio-rad obtained: back up from the investigation's database, punch file, summary and chormatograms of initial run and repeat run.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2915274-2013-00010
MDR Report Key3458984
Report Source01,07
Date Received2013-10-15
Date of Report2013-10-15
Date of Event2013-09-13
Date Mfgr Received2013-09-17
Date Added to Maude2014-01-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactSWETA PATEL
Manufacturer Street4000 ALFRED NOBEL DR.
Manufacturer CityHERCULES CA 94547
Manufacturer CountryUS
Manufacturer Postal94547
Manufacturer Phone5107415157
Single Use0
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVARIANTNBS WITH GDM 3.1 SOFTWARE
Generic NameABNORMAL HEMOGLOBIN QUANTITATION
Product CodeGKA
Date Received2013-10-15
Catalog Number250-3000
OperatorHEALTH PROFESSIONAL
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBIO-RAD LABORATORIES
Manufacturer Address4000 ALFRED NOBEL DR. HERCULES CA 94547 US 94547


Patients

Patient NumberTreatmentOutcomeDate
10 2013-10-15

© 2026 FDA.report