WARM SCALE 500111-503

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2013-10-08 for WARM SCALE 500111-503 manufactured by Natus Medical Incorporated.

Event Text Entries

[3943585] Warm-scale w/tray device was reported to have a melted tray. There was no smoke, no fire, no injuries. No patient/user involvement. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[11241134] The warm scale user manual requires annual cleaning of the heater assembly, and cleaning of the inlet filter every 6 months, or more frequently as required. Based on the amount of dirt and lint in the heater assembly, the unit had not been cleaned for some time, it is unknown when the last cleaning took place. Scale was also missing the filter, which allowed excessive dust and lint into the heater assembly and likely accelerated the problem. The manual also requires annual warmer cal. It is not known if this was done, however, the steady-state temperature from testing indicates that the unit had been changed from the factory settings. The calibration set points, as received from the customer and after cleaning the heater assembly, are about 6 degrees too high. Local temperatures at the heater outlet caused by low airflow due to obstruction are the cause of the deformed tray plastic. Additionally, the customer tray was broken in two different places which caused the tray to be misaligned with the heater assembly output vent. After performing the investigation detailed below, it is evident that the scale was not maintained according to the user manual, and as a result produced excessive temperatures at the heater output vent, which deformed the tray material.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3018859-2013-00009
MDR Report Key3459120
Report Source07
Date Received2013-10-08
Date of Report2013-10-08
Date Mfgr Received2013-09-10
Device Manufacturer Date2007-09-01
Date Added to Maude2014-01-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street5900 FIRST AVE., SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal98108
Manufacturer Phone2062685156
Single Use3
Remedial ActionIN
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameWARM SCALE
Generic NameSMART SCALE
Product CodeFRW
Date Received2013-10-08
Returned To Mfg2013-10-01
Model Number500111-503
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerNATUS MEDICAL INCORPORATED
Manufacturer AddressSEATTLE WA US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2013-10-08