MOBILEHELP CELLUAR BASE STATION (CLASSIC) IGBW-01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a * report with the FDA on 2013-11-12 for MOBILEHELP CELLUAR BASE STATION (CLASSIC) IGBW-01 manufactured by Integrity Tracking Dba Mobilehelp.

Event Text Entries

[3853393] Spoke with rich summers/son, his mother tried using system on 10/27/13 @ approx. 5pm pacific time and neck pendant would not work. His mom has a bruised knee, she was found on the floor when they were bringing her dinner.
Patient Sequence No: 1, Text Type: D, B5


[11416703] A software defect was found in this version of software on this device (r025), which causes the system to hang (rendering any future activations to not work, if the device is activated and no call back is made to the device. Two work arounds have been put in place that mitigates this issue. The monitoring center operator instructs the user to press the reset button after they hang up from the call, and the operator is instructed to answer every alarm from the device, regardless of the time between alarms (duplicate alarms). In addition a report is run by mobile help to look for situations where possible duplicate alarms have occurred, and if, found the unit is called back, which unfreezes the device. Devices with this version (r025) have been isolated to one reseller, and they have been instructed to upgrade to (r026) when the unit is returned, which fixes this issue. This issue is managed via capa (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3008101184-2013-00006
MDR Report Key3462717
Report Source*
Date Received2013-11-12
Date of Report2013-11-11
Date of Event2013-10-27
Date Added to Maude2014-02-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Single Use0
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMOBILEHELP CELLUAR BASE STATION (CLASSIC)
Generic NamePERSONAL EMERGENCY RESPONSE SYSTEM
Product CodeILQ
Date Received2013-11-12
Returned To Mfg2013-11-07
Model NumberIGBW-01
OperatorLAY USER/PATIENT
Device AvailabilityR
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerINTEGRITY TRACKING DBA MOBILEHELP
Manufacturer Address3701 FAU BLVD SUITE 300 BOCA RATON FL 33431 US 33431


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2013-11-12

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.