2.3MM 1.0MM 80CM DISPOSABLE ANGIOSCOPE ANG-080-D10K 100326502

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04,05,06,07 report with the FDA on 2013-10-11 for 2.3MM 1.0MM 80CM DISPOSABLE ANGIOSCOPE ANG-080-D10K 100326502 manufactured by Applied Medical.

Event Text Entries

[3940716] Left fem-pop bypass with insertion of saphenous vein: "disposable angioscope would not display picture during surgical procedure. "
Patient Sequence No: 1, Text Type: D, B5


[11411819] Investigation summary: the event unit was returned for eval. Upon inspection, engineering confirmed the device displayed a skewed and blurry image. No damage was noted to the unit. Engineering disassembled the unit and found that one eyepiece lens was out of its proper position. All endoscopes undergo 100% visual inspection and are functionally tested during the manufacturing process. It is likely that the lens was not secured in its proper position due to an operator error and became detached inside the eyepiece during shipping and handling of the product. The production team has been notified of this occurrence in order to develop a heightened awareness of the incident. This document represents our initial/final report.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2027111-2013-00383
MDR Report Key3466447
Report Source04,05,06,07
Date Received2013-10-11
Date of Report2013-10-09
Date of Event2013-08-09
Report Date2013-10-09
Date Reported to Mfgr2013-08-12
Date Added to Maude2013-11-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactLAUREN CONTURSI
Manufacturer Street22872 AVENIDA EMPRESA
Manufacturer CityRANCHO SANTA MARGARITA CA 92688
Manufacturer CountryUS
Manufacturer Postal92688
Manufacturer Phone9497136671
Manufacturer G1APPLIED MEDICAL
Manufacturer Street22872 AVENIDA EMPRESA
Manufacturer CityRANCHO SANTA MARGARITA CA 92688
Manufacturer CountryUS
Manufacturer Postal Code92688
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name2.3MM 1.0MM 80CM DISPOSABLE ANGIOSCOPE
Generic NameNONE
Product CodeLYK
Date Received2013-10-11
Returned To Mfg2013-08-23
Model NumberANG-080-D10K
Catalog Number100326502
Lot Number1174936
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerAPPLIED MEDICAL
Manufacturer AddressRANCHO SANTA MARGARITA CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2013-10-11

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.