MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,08 report with the FDA on 2013-11-18 for SHUNT DENVER ASCITES PAK DOUBLE VALVE 42-2050 manufactured by Carefusion.
[4023224]
The distributor reported that during incoming inspection the product had small black debris inside of the pump chamber.
Patient Sequence No: 1, Text Type: D, B5
[11336516]
(b)(4). In response to an fda inspection (conducted (b)(4) 2013), carefusion 2200 initiated a capa investigation (capa (b)(4)). As part of the capa, a retrospective review of the complaints for? Debris? And? Contamination? For applicable devices was conducted. It was determined that this report necessitates submission as a medical device report (mdr). Evaluation summary: one (1) unopened sample was provided for evaluation. Evaluation of the complaint sample confirmed the reported failure mode. A review of complaint data identified that this failure mode is not an isolated event. It has been identified that this failure mode is not an isolated event. The shunt assemblies are manufactured in a controlled manufacturing environment that uses strict control over product movement into and out of the manufacturing room and requires specialty personal protective equipment for the manufacturing personnel to minimize the inadvertent transfer of debris a device history for the listed manufacturing lot showed (8) non-conformances. In each instance, the product was 100% culled and submitted to quality for release. No other recorded quality problems or rejections were related to this incident. The manufacturing plant has also initiated a capa investigation ((b)(4)) to address the reported issue. All corrective and preventive actions (including manufacturing and quality plan improvements) will be implemented per the capa.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 1625685-2013-00071 |
MDR Report Key | 3474568 |
Report Source | 01,08 |
Date Received | 2013-11-18 |
Date of Report | 2010-08-30 |
Date of Event | 2010-08-25 |
Date Mfgr Received | 2010-11-15 |
Date Added to Maude | 2014-02-25 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | JILL RITTORNO |
Manufacturer Street | 75 NORTH FAIRWAY DRIVE |
Manufacturer City | VERNON HILLS IL 60061 |
Manufacturer Country | US |
Manufacturer Postal | 60061 |
Manufacturer Phone | 8473628056 |
Manufacturer G1 | CAREFUSION 2200, INC |
Manufacturer Street | 400 EAST FOSTER RD |
Manufacturer City | MANNFORD OK 74044 |
Manufacturer Country | US |
Manufacturer Postal Code | 74044 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | SHUNT DENVER ASCITES PAK DOUBLE VALVE |
Generic Name | SHUNT, PERITONEAL |
Product Code | KPM |
Date Received | 2013-11-18 |
Returned To Mfg | 2010-10-05 |
Model Number | 42-2050 |
Lot Number | L0J138 |
Operator | PHYSICIAN |
Device Availability | R |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | CAREFUSION |
Manufacturer Address | 75 NORTH FAIRWAY DRIVE VERNON HILLS IL 60061 US 60061 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2013-11-18 |