SHUNT DENVER ASCITES PAK DOUBLE VALVE 42-2050

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,08 report with the FDA on 2013-11-18 for SHUNT DENVER ASCITES PAK DOUBLE VALVE 42-2050 manufactured by Carefusion.

Event Text Entries

[4015555] The distributor reported that during incoming inspection two hair like fibers were noted inside the pump chamber on the shunt.
Patient Sequence No: 1, Text Type: D, B5


[11277907] (b)(4) in response to an fda inspection (conducted (b)(4) 2013), carefusion 2200 initiated a capa investigation (capa (b)(4)). As part of the capa, a retrospective review of the complaints for 'debris' and 'contamination' for applicable devices was conducted. It was determined that this report necessitates submission as a medical device report (mdr). Evaluation summary: evaluation of the sample, using 40x magnification, confirmed the presence of embedded debris in the shunt body wall. The investigation was not able to confirm the presence of debris in the fluid path. The shunt bodies are molded in a controlled manufacturing environment with strict controls on proper personal protective equipment. A device history review for the listed manufacturing lot showed (b)(4) non-conformances. In each instance, the product was 100% culled and submitted to quality for inspection and release. There were no other recorded quality problems or rejections related to this incident. The manufacturing plant has also initiated a capa investigation (b)(4) to address the reported issue. All corrective and preventive actions (including manufacturing and quality plan improvements) will be implemented per the capa.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1625685-2013-00075
MDR Report Key3474606
Report Source01,08
Date Received2013-11-18
Date of Report2010-05-19
Date of Event2010-05-19
Date Mfgr Received2010-07-14
Date Added to Maude2014-02-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJILL RITTORNO
Manufacturer Street75 NORTH FAIRWAY DRIVE
Manufacturer CityVERNON HILLS IL 60061
Manufacturer CountryUS
Manufacturer Postal60061
Manufacturer Phone8473628056
Manufacturer G1CAREFUSION 2200, INC
Manufacturer Street400 EAST FOSTER RD
Manufacturer CityMANNFORD OK 74044
Manufacturer CountryUS
Manufacturer Postal Code74044
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSHUNT DENVER ASCITES PAK DOUBLE VALVE
Generic NameSHUNT, PERITONEAL
Product CodeKPM
Date Received2013-11-18
Returned To Mfg2010-05-24
Model Number42-2050
Lot NumberL0C138
OperatorPHYSICIAN
Device AvailabilityR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCAREFUSION
Manufacturer Address75 NORTH FAIRWAY DRIVE VERNON HILLS IL 60061 US 60061


Patients

Patient NumberTreatmentOutcomeDate
10 2013-11-18

© 2024 FDA.report