TOTAL PROTEIN URINE/CSF GEN.3 03333825190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2013-11-19 for TOTAL PROTEIN URINE/CSF GEN.3 03333825190 manufactured by Roche Diagnostics.

Event Text Entries

[4018968] The customer alleged they received questionable total protein urine/csf gen. 3 (tpuc) results for one patient on their c501 analyzer. The patient's urine sample was initially tested undiluted and the result was 12 mg/dl. The initial result was reported outside the laboratory, but the physician questioned the result. The patient's sample was then diluted and the result was 14804 mg/dl. It was not confirmed whether the second result was reported to the physician. The customer also provided results from undiluted samples of 13. 2 mg/dl and 13. 2 mg/dl. It was assumed the results were from the same samples, but not confirmed. There were no adverse events. The tpuc reagent lot number was 683083 and the expiration date was 07/30/2014.
Patient Sequence No: 1, Text Type: D, B5


[11412929] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[21577753] An investigation of the reaction kinetics indicated the sample had a very high turbidity. The very high protein content in the sample interferes with the analyzer's ability to correctly interpret the concentration. This limitation is stated in the package insert.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2013-07089
MDR Report Key3476062
Report Source01,05,06
Date Received2013-11-19
Date of Report2013-12-13
Date of Event2013-11-03
Date Mfgr Received2013-11-05
Date Added to Maude2013-11-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactNA JENNIFER WOLFGRAM
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175217008
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameTOTAL PROTEIN URINE/CSF GEN.3
Generic NameTURBIDIMETRIC, TOTAL PROTEIN
Product CodeJGQ
Date Received2013-11-19
Model NumberNA
Catalog Number03333825190
Lot Number683083
ID NumberNA
Device Expiration Date2014-07-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250


Patients

Patient NumberTreatmentOutcomeDate
10 2013-11-19

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.