CASTVAC W/8 FOOT HOSE AND MOBILE STAND 0986000000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2013-11-20 for CASTVAC W/8 FOOT HOSE AND MOBILE STAND 0986000000 manufactured by Stryker Instruments-kalamazoo.

Event Text Entries

[4016666] It was reported that the castvac w/8 foot hose and mobile stand sparked and smoked while being used on a cut booth on the service production floor. There was no patient involvement and no adverse consequences associated with the device.
Patient Sequence No: 1, Text Type: D, B5


[4093061] It was reported that the castvac w/8 foot hose and mobile stand sparked and smoked while being used on a cut booth on the service production floor. There was no patient involvement and no adverse consequences associated with the device.
Patient Sequence No: 1, Text Type: D, B5


[11358856] Failure analysis is ongoing; additional information may be submitted once the results analysis is complete.
Patient Sequence No: 1, Text Type: N, H10


[11422190] The reported events, device sparked and smoke came from the unit, were confirmed. During testing the service technician confirmed the customer comments, as through troubleshooting traces of sparks and smoke were observed. The tech found the motor had evidence of sparking and overheating, and the motor was identified as the primary failure. A motor failure can cause device functionality issues along with overheating, which can lead to sparking or smoke.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number0001811755-2013-02912
MDR Report Key3478173
Report Source07
Date Received2013-11-20
Date of Report2013-10-24
Date of Event2013-10-24
Date Mfgr Received2014-01-22
Device Manufacturer Date1994-01-30
Date Added to Maude2013-11-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. CASEY METZGER
Manufacturer Street4100 EAST MILHAM AVENUE
Manufacturer CityKALAMAZOO MI 49001
Manufacturer CountryUS
Manufacturer Postal49001
Manufacturer Phone2693237700
Manufacturer G1STRYKER INSTRUMENTS-KALAMAZOO
Manufacturer Street4100 EAST MILHAM AVENUE
Manufacturer CityKALAMAZOO MI 49001
Manufacturer CountryUS
Manufacturer Postal Code49001
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCASTVAC W/8 FOOT HOSE AND MOBILE STAND
Generic NameINSTRUMENT, CAST REMOVAL, AC-POWERED
Product CodeLGH
Date Received2013-11-20
Returned To Mfg2013-10-25
Catalog Number0986000000
OperatorOTHER
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER INSTRUMENTS-KALAMAZOO
Manufacturer Address4100 EAST MILHAM AVENUE KALAMAZOO MI 49001 US 49001


Patients

Patient NumberTreatmentOutcomeDate
10 2013-11-20

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.