MOBILEHELP - CLASSIC BASE STATION IGB-01 IGB8LY22JM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04 report with the FDA on 2013-05-14 for MOBILEHELP - CLASSIC BASE STATION IGB-01 IGB8LY22JM manufactured by Integrity Tracking Llc Dba Mobilehelp.

Event Text Entries

[4020363] On (b)(6) 2013 around 10am, subscriber fell just outside of her home. Her head hit a brick and she had a gash and was bleeding badly. She said the situation was life threatening. Her daughter (b)(6) said she lives next door and heard her mom yell, when she went over to her, her mom said she already pressed the neck pendant button so subscriber assumed ems was on the way. Daughter went into the home and heard the base station beeping and called 911 and reset the base station. The mobile device was sitting in the base station charger.
Patient Sequence No: 1, Text Type: D, B5


[11256544] The device did not cause or contribute to the reported injury. The subscriber fell and pushed the button to get help. The subscribers daughter heard the mother asking for help, and called 911. The base station was beeping, and the subscribers daughter hit reset to silence the alarm. The device is intended to summons for help as needed. Device was returned and evaluated and it was determined that the base station did not connect to the monitoring center. It is unk if the mobile device connected. The device is not labeled as a life sustaining device.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3008101184-2013-00004
MDR Report Key3478968
Report Source04
Date Received2013-05-14
Date of Report2013-04-15
Date of Event2013-04-15
Date Mfgr Received2013-04-15
Device Manufacturer Date2010-12-01
Date Added to Maude2013-12-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER CAREGIVERS
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR GARY BECKER
Manufacturer Street3701 FAU BLVD SUITE 300
Manufacturer CityBOCA RATON FL 33431
Manufacturer CountryUS
Manufacturer Postal33431
Manufacturer Phone5613476255
Manufacturer G1INTEGRITY TRACKING LLC DBA MOBILEHELP
Manufacturer Street3701 FAU BLVD SUITE 300
Manufacturer CityBOCA RATON FL 33431000
Manufacturer CountryUS
Manufacturer Postal Code33431 0000
Single Use0
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMOBILEHELP - CLASSIC BASE STATION
Generic NamePERSONNAL EMERGENCY RESPONSE SYSTEM - CLASSIC, PRODUCT CODE: ILQ
Product CodeILQ
Date Received2013-05-14
Returned To Mfg2013-05-01
Model NumberIGB-01
Catalog NumberIGB8LY22JM
OperatorOTHER
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerINTEGRITY TRACKING LLC DBA MOBILEHELP
Manufacturer Address3701 FAU BLVD SUITE 300 BOCA RATON FL 33431 US 33431


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2013-05-14