PERIOSTEAL ELEVATOR 6MM CURVED BLADE-ROUND EDGE 399.36

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,07 report with the FDA on 2013-11-20 for PERIOSTEAL ELEVATOR 6MM CURVED BLADE-ROUND EDGE 399.36 manufactured by Synthes (usa).

Event Text Entries

[18311519] On (b)(6) 2013, a drill sleeve became stuck with a schanz screw while an external fixation was placed on a male patient. The sleeve and screw were not able to be separated from each other. It was reported that the shaft of the screw appeared to be bent. A larger sleeve with another screw was used to complete the procedure. It was reported that the procedure was delayed six minutes and was successfully completed. The reporter tried unsuccessfully to separate the sleeve and screw on the back table using a periosteal elevator (399. 36) and reduction forceps. In the attempt, the tip of the periosteal elevator and the handle of the reduction forceps were broken. This is report 1 of 1 for complaint (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[18398305] (b)(6). Device is an instrument and is not implanted/explanted. Subject device has been received and is currently in the evaluation process. Investigation is on going; no conclusion could be drawn. Placeholder.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2520274-2013-07460
MDR Report Key3479058
Report Source05,07
Date Received2013-11-20
Date of Report2013-10-23
Date of Event2013-10-23
Date Mfgr Received2013-10-23
Date Added to Maude2014-02-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactELIZABETH FIGLEAR
Manufacturer Street1302 WRIGHTS LANE EAST
Manufacturer CityWEST CHESTER PA 19380
Manufacturer CountryUS
Manufacturer Postal19380
Manufacturer Phone8006207025
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePERIOSTEAL ELEVATOR 6MM CURVED BLADE-ROUND EDGE
Product CodeHTE
Date Received2013-11-20
Returned To Mfg2013-11-06
Catalog Number399.36
Lot Number6022
OperatorLAY USER/PATIENT
Device AvailabilityR
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSYNTHES (USA)
Manufacturer Address1302 WRIGHTS LANE EAST WEST CHESTER PA 19380 US 19380


Patients

Patient NumberTreatmentOutcomeDate
10 2013-11-20

© 2026 FDA.report