ARCHITECT CYCLOSPORINE 01L75-25

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05 report with the FDA on 2013-11-26 for ARCHITECT CYCLOSPORINE 01L75-25 manufactured by Abbott Laboratories.

Event Text Entries

[4001719] The customer observed a falsely elevated cyclosporine result for one patient on the architect i2000sr analyzer. The following data was provided: sid (b)(6) initial 992. 5 ng/ml, repeat with a new pretreatment 31. 9 ng/ml. The customer states that based on patient history the second result is correct as the patient is under therapy for post-transplant. There was no impact to patient management reported.
Patient Sequence No: 1, Text Type: D, B5


[11358308] An evaluation is in process.
Patient Sequence No: 1, Text Type: N, H10


[26632316] Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, a review of labeling and accuracy testing. No adverse trend was identified for the customer's issue. Labeling was reviewed and found to be adequate. All testing met all specifications. Complaint information reasonably suggests the assay is performing as intended, and no malfunction of the device occurred. Based on the available information, no product deficiency was identified.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1415939-2013-00403
MDR Report Key3490291
Report Source01,05
Date Received2013-11-26
Date of Report2013-11-06
Date of Event2013-11-06
Date Mfgr Received2014-02-18
Device Manufacturer Date2013-04-01
Date Added to Maude2013-12-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone847937-512
Manufacturer G1ABBOTT LABORATORIES
Manufacturer Street100 ABBOTT PARK ROAD
Manufacturer CityABBOTT PARK IL 60064350
Manufacturer CountryUS
Manufacturer Postal Code60064 3500
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameARCHITECT CYCLOSPORINE
Product CodeMKW
Date Received2013-11-26
Catalog Number01L75-25
Lot Number24427M500
OperatorOTHER
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT LABORATORIES
Manufacturer Address100 ABBOTT PARK ROAD ABBOTT PARK IL 60064350 US 60064 3500


Patients

Patient NumberTreatmentOutcomeDate
10 2013-11-26

© 2024 FDA.report