AU681 CLINICAL CHEMISTRY ANALYZER AU681-02E, CHEMISTRY ANALYZER AU680 WITH ISE, 210V N3147200

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2013-12-03 for AU681 CLINICAL CHEMISTRY ANALYZER AU681-02E, CHEMISTRY ANALYZER AU680 WITH ISE, 210V N3147200 manufactured by Beckman Coulter.

Event Text Entries

[19968438] The customer experienced erratic potassium results on the ise side of their au681 instrument. The customer got a result of 6. 6, which, when repeat, resulted in a value of 4. 4. The day before, a result of 7. 6 was obtained and when repeated it was 4. 0. Erroneous results were generated but not reported outside the laboratory, so there was no effect to patient or change to patient treatment. During troubleshooting, the customer observed bubbles within the reference line.
Patient Sequence No: 1, Text Type: D, B5


[20481823] A field service engineer (fse) was dispatched and upon arrival verified customer's issue. He replaced the reference( ref) valve. He also replaced the peristaltic pump tubings and all related ise tubings. He then calibrated the ise three times and ran qc. All results were within specifications. The instrument was operational. The manufacturer's reference # for this event is (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[34092161] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9612296-2013-00168
MDR Report Key3496095
Report Source05,06
Date Received2013-12-03
Date of Report2013-11-11
Date of Event2013-11-10
Date Mfgr Received2013-11-11
Device Manufacturer Date2008-01-11
Date Added to Maude2014-03-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. DUNG NGUYEN
Manufacturer Street250 S. KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149614941
Manufacturer G1BECKMAN COULTER MISHIMA K.K.
Manufacturer Street454-32 HIGASHINO NAGAIZUMI-CHO
Manufacturer CitySUNTO-GUN, SCHIZUOKA 411-093
Manufacturer CountryJA
Manufacturer Postal Code411-093
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAU681 CLINICAL CHEMISTRY ANALYZER
Generic NameANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE
Product CodeMZV
Date Received2013-12-03
Model NumberAU681-02E, CHEMISTRY ANALYZER AU680 WITH ISE, 210V
Catalog NumberN3147200
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER
Manufacturer Address250 S. KRAEMER BLVD. BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2013-12-03

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.