AU680 CLINICAL CHEMISTRY ANALYZER B12188

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2013-12-05 for AU680 CLINICAL CHEMISTRY ANALYZER B12188 manufactured by Beckman Coulter.

Event Text Entries

[4145617] A customer reported erroneous low potassium test results were obtained for one patient sample when using the au680 clinical chemistry analyzer. The customer stated that these results were not reported outside of the laboratory. There was no death, injury or affect to patient treatment.
Patient Sequence No: 1, Text Type: D, B5


[11271114] The customer observed bubbling within the reference block and were unsure of the source. The field service engineer (fse) evaluated the analyzer and verified the customer's observation. The fse replaced the ref (reference) valve and ise (ion selective electrode) tubing as a preventative measure. The fse then primed the ise and did not observe any bubbles. Calibrations, precision and qc were performed, with all results passing within specifications. Identification of failure mode: ref valve was replaced and bubbles were removed within the line.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9612296-2013-00171
MDR Report Key3501945
Report Source05,06
Date Received2013-12-05
Date of Report2013-11-15
Date of Event2013-11-15
Date Mfgr Received2013-11-15
Device Manufacturer Date2013-07-24
Date Added to Maude2014-03-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. DUNG NGUYEN
Manufacturer Street250 S KRAEMER BLVD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149614941
Manufacturer G1BECKMAN COULTER MISHIMA K.K.
Manufacturer Street454-32 HIGASHINO NAGAIZUMI-CHO
Manufacturer CitySUNTO-GUN, SHIZUOKA, P-NOTA 411-0931
Manufacturer CountryJA
Manufacturer Postal Code411-0931
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameAU680 CLINICAL CHEMISTRY ANALYZER
Generic NameANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE
Product CodeMZV
Date Received2013-12-05
Model NumberNA
Catalog NumberB12188
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER
Manufacturer Address250 S. KRAEMER BLVD. BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2013-12-05

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.