TVC INSIGHT CATHETER TVC-C195-20

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2013-11-19 for TVC INSIGHT CATHETER TVC-C195-20 manufactured by Infraredx Inc..

Event Text Entries

[19199730] Pt presented for pci. During procedure, infraredx catheter was prepped according to ifu. It was inserted into the om2 and at that time the fellow noticed the pt had no pressure. Doctor tightened the touhy and still no pressure on the monitor. He decided to proceed with pullback. The results were outliers, indicative of air in either the catheter or vessel. The pt began crashing at this time and the catheter was removed. After intubation and cpr, doctor took a quick angiogram and the vessel was closed. Doctor intervened, using an export catheter and medication. Pt stabilized. Doctor confirmed vessel was open. Doctor believes root cause to be an air embolus and not catheter.
Patient Sequence No: 1, Text Type: D, B5


[19389524] Catheter was returned to manufacturer for investigation. Upon receipt, catheter was inspected for any visual or manufacturing defects. None found. Catheter was then tested on a test bed and found to be fully functional, producing functional chemograms and ivus info.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004722468-2013-00004
MDR Report Key3503544
Report Source07
Date Received2013-11-19
Date of Report2013-11-16
Date of Event2013-10-16
Date Mfgr Received2013-10-17
Date Added to Maude2013-12-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactSTEVE CHARTIER
Manufacturer Street34 THIRD AVENUE
Manufacturer CityBURLINGTON MA 01803
Manufacturer CountryUS
Manufacturer Postal01803
Manufacturer Phone7813459593
Single Use0
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTVC INSIGHT CATHETER
Generic NameNIR-IVUS CORONARY IMAGING CATHETER
Product CodeOGZ
Date Received2013-11-19
Model NumberTVC-C195-20
Lot Number304242-A
Device Expiration Date2014-12-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerINFRAREDX INC.
Manufacturer Address34 THIRD AVENUE BURLINGTON


Patients

Patient NumberTreatmentOutcomeDate
101. Life Threatening; 2. Required No Informationntervention 2013-11-19

© 2026 FDA.report