OPSIS Z-CAM VS7100 *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 1996-08-07 for OPSIS Z-CAM VS7100 * manufactured by V. Mueller Div. Baxther Healthcare Corp..

Event Text Entries

[22430] Surgeons complained the camera was blurry and out of focus, and unable to see anatomy. Laparoscopic cholecystectomy converted to open procedure. Biomedical engineering tech evaluation of camera system found that the camera picture has too much chroma (purple). During the past year the camera has "hsd" problems with color changing and an inability to maintain color during surgery cases. The camera has also been fluctuating in light intensity, which has made it difficult to distinguish the color of tissue. The hosp owns 2 camera systems and both have a history of poor quality picture. Both have been returned to the mfr twice during the past year for various problems, the mfr has not found any problem with the color fluctuations with the cameras. The monitors have also been returned to the mfr for evaluation of color changes. Replaced a diode in each to improve the picture contrast.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW1009658
MDR Report Key35041
Date Received1996-08-07
Date of Report1996-07-29
Date of Event1996-07-11
Date Added to Maude1996-08-09
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA0
Report to FDA0
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameOPSIS Z-CAM
Generic NameSURGICAL CAMERA BOX
Product CodeFWC
Date Received1996-08-07
Model NumberVS7100
Catalog Number*
Lot Number*
ID NumberBC#70974
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key36458
ManufacturerV. MUELLER DIV. BAXTHER HEALTHCARE CORP.
Manufacturer Address1500 WAUKEGAN RD MCGAW PARK IL 60085 US


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 1996-08-07

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.