FREEDOM EVO-2 200 MCA 30020020

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 99 report with the FDA on 2013-11-19 for FREEDOM EVO-2 200 MCA 30020020 manufactured by Tecan Schweiz.

Event Text Entries

[3963895] Laboratory called tecan on (b)(6) to report that during a test run of an evo 200 mca liquid handling pipettor instrument, the operator had pressed the pause button and reached inside the instrument to remove tips that were stuck to the mca head. The pause button was lit indicating instrument was paused. While the operator's hand was inside the instrument, the mca head moved down pinning the operator's hand to the deck of the instrument. The operator hit the pause button to unpause the instrument and the mca head moved up freeing the operator's hand. The operator sustained minor injury to the hand including a cut with bleeding. No medical treatment was required other than basic first aid. The test run was without reagents and the tips were clean and sterile. No risk of biohazard exposure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1037985-2013-00002
MDR Report Key3504644
Report Source99
Date Received2013-11-19
Date of Report2013-11-15
Date of Event2013-11-01
Date Facility Aware2013-11-01
Report Date2013-11-15
Date Reported to FDA2013-11-18
Date Reported to Mfgr2013-11-04
Date Added to Maude2013-12-11
Event Key0
Report Source CodeDistributor report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street4022 STIRRUP CREEK RD.
Manufacturer CityRESEARCH TRIANGLE PARK NC 27709
Manufacturer CountryUS
Manufacturer Postal27709
Manufacturer G1TECAN U.S.
Manufacturer Street4022 STIRRUP CREEK RD.
Manufacturer CityRESEARCH TRIANGLE PARK NC 27709
Manufacturer CountryUS
Manufacturer Postal Code27709
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameFREEDOM EVO-2 200 MCA
Generic NamePIPETTING STATION FOR CLINICAL USE
Product CodeJQW
Date Received2013-11-19
Catalog Number30020020
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age3 YR
Device Sequence No1
Device Event Key0
ManufacturerTECAN SCHWEIZ
Manufacturer AddressMAENNEDORF ZUERICH SZ


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2013-11-19

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.