RELIANCE 220/220L CART WASHER

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2013-12-13 for RELIANCE 220/220L CART WASHER manufactured by Steris Canada Corporation.

Event Text Entries

[3965709] The user facility reported their unit was leaking steam. The fire alarm and sprinkler systems were not activated and the facility did not require evacuation. No injuries or procedural delays/cancellations were reported.
Patient Sequence No: 1, Text Type: D, B5


[11324676] A steris service technician arrived at the facility, inspected the washer and found the hot water tank was overheating. The technician also observed the solenoid valve below the unit's hot water tank was leaking water. The technician replaced the temperature control thermostat which was causing the tank to overheat and replaced the solenoid valve. The thermostat was properly adjusted, a test cycle was completed and the unit was returned to operation. The following day, the user facility reported the unit did not have power. The steris service technician returned to the facility and found the hot water fill valve flare fitting was leaking water onto the unit's thermostat, causing a short of the thermostat and a constantly energized steam valve. The energized steam valve severely overheated the unit's hot water tank and caused water to boil over the tank and down onto the unit's electrical box causing the unit to shut down. The technician replaced the damaged thermostat and solenoid valve, tightened the flare fitting, dried the electrical box, ran a test cycle and returned the unit to service. The washer was installed in june of 1998 and is currently under a steris service contract. The last preventive maintenance was performed on october 23, 2013 at which time the unit was confirmed to be operating to specification.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9680353-2013-00164
MDR Report Key3519306
Report Source06
Date Received2013-12-13
Date of Report2013-12-13
Date of Event2013-11-25
Date Mfgr Received2013-11-25
Date Added to Maude2014-03-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS KATHRYN CADORETTE
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927231
Manufacturer G1STERIS CANADA CORPORATION
Manufacturer Street490, ARMAND-PARIS
Manufacturer CityQUEBEC, QUEBEC GIC8A3
Manufacturer CountryCA
Manufacturer Postal CodeGIC 8A3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRELIANCE 220/220L CART WASHER
Generic NameWASHER
Product CodeFLH
Date Received2013-12-13
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTERIS CANADA CORPORATION
Manufacturer Address490 ARMAND-PARIS QUEBEC, QUEBEC GIC8A3 CA GIC 8A3


Patients

Patient NumberTreatmentOutcomeDate
10 2013-12-13

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.