INGENUITY TF PET/MR 882380

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06,07 report with the FDA on 2013-12-11 for INGENUITY TF PET/MR 882380 manufactured by Philips Medical Systems(cleveland), Inc..

Event Text Entries

[4128433] The customer reported a left to right misalignment with pet reconstructed images. The images appear diagnostic and show a finding that is typical with patients with memory disorder. After reconstructing data again, the finding disappears. Therefore, the initial reconstruction has left to false positive results in more than one patient. The issue was discovered during a research protocol, when the scanning of healthy volunteers presented with the same findings as those patients with known memory disorders. The healthy volunteers have undergone a series of neuropsychological tests and have been determined healthy. The discrepancy between the imaging findings and neuropsychological results has led to further investigation and the site is currently reviewing patient data.
Patient Sequence No: 1, Text Type: D, B5


[11458289] Note: we have not completed our investigation of this event. We will file a follow-up mdr at the completion of the investigation. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1525965-2013-00440
MDR Report Key3525443
Report Source05,06,07
Date Received2013-12-11
Date of Report2013-11-25
Date Mfgr Received2013-11-25
Date Added to Maude2014-03-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactKUMUDINI CARTER
Manufacturer Street595 MINER RD.
Manufacturer CityCLEVELAND OH 44143
Manufacturer CountryUS
Manufacturer Postal44143
Manufacturer Phone4404833032
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameINGENUITY TF PET/MR
Product CodeOUO
Date Received2013-12-11
Model Number882380
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerPHILIPS MEDICAL SYSTEMS(CLEVELAND), INC.
Manufacturer Address595 MINER RD. CLEVELAND OH 44143 US 44143


Patients

Patient NumberTreatmentOutcomeDate
10 2013-12-11

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.