ADVIA CENTAUR XP FSH N/A 01360521

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2013-12-20 for ADVIA CENTAUR XP FSH N/A 01360521 manufactured by Siemens Healthcare Diagnostics, Inc..

Event Text Entries

[19588119] Customer performed a method comparison with 75 samples between advia centaur xp fsh reagent lots 033186 and 033188. The samples tested were across the range of the assay (0. 3 to 200 miu/ml). An average positive bias of 17% was observed by the customer with reagent lot 033188 compared to reagent lot 033186. There were no reports that treatment was altered or prescribed based on these results. There are no reports of adverse health consequences based on these results.
Patient Sequence No: 1, Text Type: D, B5


[19824045] Siemens performed an internal study comparing advia centaur fsh reagent lot 033186 and 033188. Quality control material results were all within range. Patient samples did not show the consistent bias observed by the customer. The customer gathered additional data and determined that the overall patient bias that they were observing is 9%. The root cause of the initial positive bias with advia centaur xp fsh reagent lot 033188 cannot be determined based on the information available.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1219913-2013-00312
MDR Report Key3531209
Report Source05,06
Date Received2013-12-20
Date of Report2013-11-25
Date of Event2013-11-16
Date Mfgr Received2013-11-25
Date Added to Maude2014-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactLOUISE MCLAUGHLIN
Manufacturer Street333 CONEY STREET
Manufacturer CityEAST WALPOLE MA 02032
Manufacturer CountryUS
Manufacturer Postal02032
Manufacturer Phone5086604381
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS, INC
Manufacturer Street333 CONEY STREET
Manufacturer CityEAST WALPOLE MA 02032
Manufacturer CountryUS
Manufacturer Postal Code02032
Single Use3
Remedial ActionIN
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameADVIA CENTAUR XP FSH
Generic NameRADIOIMMUNOASSAY, FOLLICLE-STIMULATING HORMONE
Product CodeCGJ
Date Received2013-12-20
Model NumberN/A
Catalog Number01360521
Lot Number033188
Device Expiration Date2014-05-26
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS, INC.
Manufacturer AddressTARRYTOWN NY 10591 US 10591


Patients

Patient NumberTreatmentOutcomeDate
10 2013-12-20