MOBILEHELP BASE STATION - IGB-01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04 report with the FDA on 2013-12-23 for MOBILEHELP BASE STATION - IGB-01 manufactured by Integrity Tracking Llc / Dba Mobilehelp.

Event Text Entries

[4127085] Spoke with (b)(6) who reported that his wife, (b)(6) had pressed her black wrist button (bwb) on saturday at 4:45 pm cst (b)(6) 2013 and the led flashed but the base station(bs) did not respond, classic system. Mr. (b)(6) was with his wife at that time when she was having a seizure in the living room so he pressed that black wrist button she was wearing but it flashed and did not activate the alarm, the base station was in the same room. He called 911 directly and she required transport to the hospital.
Patient Sequence No: 1, Text Type: D, B5


[11387855] The pendant lost its training to the base station igb-01. The pendant was retrained to the unit by the customer support tech. The system was then tested and worked as designed. The subscriber did not want to replace the system. The subscriber was sent the summer newsletter on 6/20/13, which included instructions on pressing the button the correct way to avoid loss of training from occurring. This issue is being managed in capa (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3008101184-2013-00007
MDR Report Key3533683
Report Source04
Date Received2013-12-23
Date of Report2013-12-23
Date of Event2013-11-09
Date Mfgr Received2013-11-09
Date Added to Maude2014-04-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER CAREGIVERS
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. GARY BECKER
Manufacturer Street3701 FAU BLVD SUITE 300
Manufacturer CityBOCA RATON FL 33431
Manufacturer CountryUS
Manufacturer Postal33431
Manufacturer Phone5613476255
Single Use3
Remedial ActionMA
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMOBILEHELP BASE STATION - IGB-01
Generic NamePERSONAL EMERGENCY RESPONSE SYSTEM
Product CodeILQ
Date Received2013-12-23
Model NumberIGB-01
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerINTEGRITY TRACKING LLC / DBA MOBILEHELP
Manufacturer Address3701 FAU BLVD SUITE 300 BOCA RATON FL 33431 US 33431


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2013-12-23

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.