MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2001-10-03 for TI 100 ROADRUNNER INFANT INCUBATOR manufactured by Hill-rom/ Air Shields.
[209007]
Roadrunner incubator stopped working and developed a burning smell. Unit had bad battery and power supply. Replaced both and reported to mfr.
Patient Sequence No: 1, Text Type: D, B5
Report Number | MW1022998 |
MDR Report Key | 354589 |
Date Received | 2001-10-03 |
Date of Report | 2001-09-28 |
Date of Event | 2001-08-13 |
Date Added to Maude | 2001-10-09 |
Event Key | 0 |
Report Source Code | Voluntary report |
Manufacturer Link | N |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 0 |
Reporter Occupation | BIOMEDICAL ENGINEER |
Health Professional | 3 |
Initial Report to FDA | 0 |
Report to FDA | 0 |
Event Location | 3 |
Single Use | 0 |
Previous Use Code | 0 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | TI 100 ROADRUNNER INFANT INCUBATOR |
Generic Name | TRANSPORT INCUBATOR |
Product Code | FPL |
Date Received | 2001-10-03 |
Model Number | TI 100 ROADRUNNER |
Catalog Number | TI 100 |
Lot Number | NA |
ID Number | * |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Implant Flag | N |
Date Removed | A |
Device Sequence No | 1 |
Device Event Key | 343809 |
Manufacturer | HILL-ROM/ AIR SHIELDS |
Manufacturer Address | 330 JACKSONVILLE RD HATBORO PA 190402211 US |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Other | 2001-10-03 |