SYNERGEYES HD HYBRID CONTACT LENSES SEE H10 NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2013-12-26 for SYNERGEYES HD HYBRID CONTACT LENSES SEE H10 NA manufactured by Synergeyes, Inc..

Event Text Entries

[4049979] On (b)(6) 2013, synergeyes received a complaint wherein the practitioner stated that the patient's poorly fit lens caused a corneal erosion which led to a corneal abrasion the patient was treated with besivance tid and lotemax qid. The patient stopped wearing the lenses and the signs and symptoms of the corneal abrasion rapidly resolved. Synergeyes suggested that the practitioner refit the lenses.
Patient Sequence No: 1, Text Type: D, B5


[11495975] There were two lenses involved in the event: right eye (od) and left eye (os). Od lens - model number: mp7787-0225+017519, lot number: 048792, expiration date: 12/01/2017, date manufactured: 01/01/2013. Os lens - model number: mp7787-0200+017522, lot number: 048573, expiration date: 12/01/2017, date manufactured: 01/01/2013. The lenses were measured for base curve and power. The results demonstrated that the lens met the labeled parameters. Results of the surface inspection determined that there were no defects on the front or base curve surfaces.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3005087645-2013-00016
MDR Report Key3553312
Report Source05
Date Received2013-12-26
Date of Report2013-12-26
Date of Event2013-09-20
Date Mfgr Received2013-12-03
Date Added to Maude2014-01-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactKAREN KINCADE, DIRECTOR
Manufacturer Street2232 RUTHERFORD ROAD
Manufacturer CityCARLSBAD CA 92008
Manufacturer CountryUS
Manufacturer Postal92008
Manufacturer Phone7604449636
Manufacturer G1SYNERGEYES, INC.
Manufacturer Street2232 RUTHERFORD ROAD
Manufacturer CityCARLSBAD CA 92008
Manufacturer CountryUS
Manufacturer Postal Code92008
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSYNERGEYES HD HYBRID CONTACT LENSES
Generic NameCONTACT LENSES
Product CodeHQD
Date Received2013-12-26
Returned To Mfg2013-12-03
Model NumberSEE H10
Catalog NumberNA
Lot NumberSEE H 10
ID NumberNA
OperatorLAY USER/PATIENT
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSYNERGEYES, INC.
Manufacturer AddressCARLSBAD CA US


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2013-12-26