ENOVATE CART EMC T.CD CART WITH L.I PHOA BATTERY, ELECTRONIC LIFT ACTUATOR AND C HCO 14458

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2013-12-31 for ENOVATE CART EMC T.CD CART WITH L.I PHOA BATTERY, ELECTRONIC LIFT ACTUATOR AND C HCO 14458 manufactured by .

Event Text Entries

[4219143] On this date an equipment cart that has a battery pack was found smoking (the battery pack) by a nurse and the staff peppered it with a fire extinguisher to stop the smoking. This was treated as a potential fire situation on the unit. A enovate care serial # (b)(4) with a li-phoa battery, electronic lift actuator and configured with medication drawers was involved in the event. No damages or injuries but the area required major clean up due to extensive fire extinguisher residue and pts were relocated to another unit. The cart was used to assist staff in dispensing medications and pt charting. The cart was pulled the services and wiring/cables in battery compartment were found to be melted. All other carts in the facility were systematic checked for the same problem and only one other cart was found to have a similar problem (hco # 14829) and both pulled from manufacturer. (http://www. Enovatemedical. Com/support. Html). The devices were returned to the vendor engineers to investigate the cause and determine if any solutions are available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5033713
MDR Report Key3556404
Date Received2013-12-31
Date of Report2013-12-31
Date of Event2013-11-07
Date Added to Maude2014-01-08
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameENOVATE CART EMC T.CD CART WITH L.I PHOA BATTERY, ELECTRONIC LIFT ACTUATOR AND C
Generic NameWHEELED CART
Product CodeBZN
Date Received2013-12-31
Returned To Mfg2013-11-14
Model NumberHCO 14458
Lot NumberPO # 9011062
Device Expiration Date2011-02-21
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Sequence No1
Device Event Key0

Device Sequence Number: 2

Brand NameENOVATE CART EMC T.CD CART WITH L.I PHOA BATTERY, ELECTRONIC LIFT ACTUATOR AND C
Generic NameWHEELED CART
Product CodeBZN
Date Received2013-12-31
Returned To Mfg2013-11-14
Model NumberHCO 14829
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Sequence No2
Device Event Key0


Patients

Patient NumberTreatmentOutcomeDate
10 2013-12-31

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.