Event Text Entries

[4074035] At the start of the laparoscopic appendectomy, as usual, the surgeon cleaned the opening of the trocar with the cotton tipped applicator to remove blood for a clearer visual. The cotton tip slipped off of the wooden handle and fell inside the abdomen. The surgeon successfully retrieved it. It does not appear as if the glue bonding the cotton tip to the wooden handle is strong enough to keep it intact. The staff has tested several applicators from several lot numbers including the ones put into our packs and they are all equally easily removed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3556546
MDR Report Key3556546
Date Received2013-11-26
Date of Report2013-11-26
Date of Event2013-10-22
Report Date2013-11-26
Date Reported to FDA2013-11-26
Date Reported to Mfgr2014-01-07
Date Added to Maude2014-01-07
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand Name*
Generic NameCOTTON TIPPED APPLICATOR
Product CodeKXG
Date Received2013-11-26
Model NumberMDS202000
Catalog Number*
Lot Number*
ID Number*
OperatorPHYSICIAN
Device AvailabilityN
Device Sequence No1
Device Event Key0
ManufacturerMEDLINE INDUSTRIES, INC.
Manufacturer AddressONE MEDLINE PLACE MUNDELEIN, IL 60060 IL 60060 US 60060


Patients

Patient NumberTreatmentOutcomeDate
10 2013-11-26

© 2026 FDA.report