HEMOCHRON JR. CITRATE APTT CUVETTE J103C

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2013-11-05 for HEMOCHRON JR. CITRATE APTT CUVETTE J103C manufactured by International Technidyne Corp..

Event Text Entries

[4048002] Healthcare professional reports higher than expected results with hemochron jr. Citrated aptt cuvette assay. During pre-operative screening for an unspecified procedure, the aptt test generated result of 72 plasma equivalent second and a repeated test generated 74 plasma equivalent seconds, which were both higher than expected. The healthcare facility's expected normal range is 21-35 seconds. A sample was tested at the reference laboratory for confirmation and results was 36 seconds. The patient was evaluated based on the laboratory result. No adverse event(s) reported.
Patient Sequence No: 1, Text Type: D, B5


[11497297] (b)(4). Method: actual device not evaluated (cuvette/single-use disposable). Result: results pending completion of evaluation. Conclusion: conclusion not yet available - evaluation in progress. Itc has requested all date required for form 3500a.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2248721-2013-00060
MDR Report Key3561106
Report Source06
Date Received2013-11-05
Date of Report2013-10-08
Date of Event2013-10-01
Date Mfgr Received2013-10-08
Device Manufacturer Date2013-05-01
Date Added to Maude2014-02-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactELEANOR FOX
Manufacturer Street20 CORPORATE PLACE S
Manufacturer CityPISCATAWAY NJ 08854
Manufacturer CountryUS
Manufacturer Postal08854
Manufacturer Phone7325485700
Manufacturer Street6-8 OLSEN AVE.
Manufacturer CityEDISON NJ 08820
Manufacturer CountryUS
Manufacturer Postal Code08820
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHEMOCHRON JR. CITRATE APTT CUVETTE
Product CodeGFO
Date Received2013-11-05
Catalog NumberJ103C
Lot NumberE3JCC013-P1
Device Expiration Date2014-05-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerINTERNATIONAL TECHNIDYNE CORP.
Manufacturer AddressEDISON NJ US


Patients

Patient NumberTreatmentOutcomeDate
10 2013-11-05