SOFT IMPACT MALLET W/ EASY GRIP HANDLES 7832

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 99 report with the FDA on 2014-01-02 for SOFT IMPACT MALLET W/ EASY GRIP HANDLES 7832 manufactured by Medtorque Inc..

Event Text Entries

[4059530] While in use during surgery, the delrin head of the mallet broke off. The delrin head was found and the device sequestered. No injuries were sustained by the patient. The user facility was informed that the part number had undergone a design change to allow enhanced immobility upon impact. To accomplish this, the re-design incorporated a metal insert within the threads of the heads to create a more tight grip. Note that the design change was already in process when this event occurred. The user facility was issued two loaner devices to use until the design change process was complete. After new devices were manufactured, replacements will be sent to the customer. A follow up message was left regarding the replacement devices.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1833053-2013-00009
MDR Report Key3572125
Report Source99
Date Received2014-01-02
Date of Report2013-12-26
Date of Event2013-10-04
Date Facility Aware2013-10-09
Report Date2013-12-26
Date Reported to FDA2013-12-26
Date Reported to Mfgr2013-12-26
Date Added to Maude2014-01-14
Event Key0
Report Source CodeDistributor report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameSOFT IMPACT MALLET W/ EASY GRIP HANDLES
Product CodeHXL
Date Received2014-01-02
Returned To Mfg2013-10-13
Model Number7832
Catalog Number7832
Lot Number1012MT
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age2 YR
Device Sequence No1
Device Event Key0
ManufacturerMEDTORQUE INC.
Manufacturer AddressKENOSHA WI US


Patients

Patient NumberTreatmentOutcomeDate
10 2014-01-02

© 2026 FDA.report