HEMOCHRON JR. CITRATE APTT CUVETTE J103C

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2013-11-14 for HEMOCHRON JR. CITRATE APTT CUVETTE J103C manufactured by International Technidyne Corp.

Event Text Entries

[17385348] Healthcare professional reports inconsistent results and results higher than reference with the hemochron jr. Citrated aptt cuvette assay. Pt being evaluated for ischemic stroke with symptoms of right side facial droop and weakness on right side. The healthcare facility's criteria to administer tissue plasminogen activator (tpa) is in part based on aptt results targeted at greater than or equal 38. 6 seconds. Repeated hemochron jr. Citrated aptt tests generated results of 46. 8, 39. 3, and 38. 3 plasma equivalent seconds. Sample sent to the reference laboratory for result confirmation generated 30. 7 seconds. The pt was evaluated and treated based on the laboratory result. No adverse event(s) reported.
Patient Sequence No: 1, Text Type: D, B5


[17712850] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2248721-2013-00063
MDR Report Key3577043
Report Source06
Date Received2013-11-14
Date of Report2013-10-21
Date of Event2013-10-01
Date Mfgr Received2013-10-21
Device Manufacturer Date2013-06-01
Date Added to Maude2014-02-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactELEANOR FOX
Manufacturer Street20 CORPORATE PLACE SOUTH
Manufacturer CityPISCATAWAY NJ 08854
Manufacturer CountryUS
Manufacturer Postal08854
Manufacturer Phone7325485700
Manufacturer Street8 OLSEN AVE
Manufacturer CityEDISON NJ 08820
Manufacturer CountryUS
Manufacturer Postal Code08820
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHEMOCHRON JR. CITRATE APTT CUVETTE
Product CodeGFO
Date Received2013-11-14
Returned To Mfg2013-10-25
Catalog NumberJ103C
Lot NumberF3JCC015-P3
Device Expiration Date2014-06-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerINTERNATIONAL TECHNIDYNE CORP
Manufacturer AddressEDISON NJ 08820 US 08820


Patients

Patient NumberTreatmentOutcomeDate
10 2013-11-14

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.