IMMAGE? 800 IMMUNOCHEMISTRY SYSTEM A15445

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2014-01-28 for IMMAGE? 800 IMMUNOCHEMISTRY SYSTEM A15445 manufactured by Beckman Coulter.

Event Text Entries

[4095095] A customer stated that the sample syringe was cracked and leaking on top of their image 800 instrument. The leaking fluid that the customer found was diluted patient sample. No erroneous patient results were generated due to this event. In addition, no biohazard exposure or injury occurred due to this event.
Patient Sequence No: 1, Text Type: D, B5


[11540722] A field service engineer (fse) called the customer and determined that the syringe barrel was completely snapped out from its fitting, causing the leak. The service engineer sent a new barrel to the customer to resolve the issue. On the following day, the service engineer called the customer and confirmed that the instrument qc was running without any further issue. This failure mode has a potential to impact patient results. The manufacturer reference number for this event is (b)(4). Customer resolved the issue.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2014-00039
MDR Report Key3596192
Report Source05,06
Date Received2014-01-28
Date of Report2014-01-09
Date of Event2014-01-09
Date Mfgr Received2014-01-09
Device Manufacturer Date2010-09-13
Date Added to Maude2014-04-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. DUNG NGUYEN
Manufacturer Street250 S. KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149614941
Manufacturer G1BECKMAN COULTER
Manufacturer Street250 S. KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal Code92821
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIMMAGE? 800 IMMUNOCHEMISTRY SYSTEM
Generic NameNEPHELOMETER, FOR CLINICAL USE
Product CodeJQX
Date Received2014-01-28
Model NumberNA
Catalog NumberA15445
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER
Manufacturer Address250 S. KRAEMER BLVD. BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2014-01-28

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.