GROCOTTS METHENAMINE SILVER STAIN * S7440-8 (87008)

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2001-11-05 for GROCOTTS METHENAMINE SILVER STAIN * S7440-8 (87008) manufactured by Richard-allan Scientific.

Event Text Entries

[18825894] Bronchial fluid sent to cytology was contaminated during processing with fungal elements. Pt results were released: stating "positive for fungus". The next day, another specimen had the same fungal elements, the slides were compared and similarities lead to the conclusion that this was a contaminant. Through the next several weeks, a thorough investigation as to the source of the contaminant was conducted. Eventually it was discovered that a purchased stain product was being delivered to hospital already contaminated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number359816
MDR Report Key359816
Date Received2001-11-05
Date of Report2001-11-05
Date of Event2001-06-21
Date Facility Aware2001-06-21
Report Date2001-11-05
Date Reported to FDA2001-11-05
Date Reported to Mfgr2001-09-21
Date Added to Maude2001-11-15
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameGROCOTTS METHENAMINE SILVER STAIN
Generic NameFAST GREEN STAIN USED TO PROCESS CYTOLOGY SPECIMENS
Product CodeHYC
Date Received2001-11-05
Model Number*
Catalog NumberS7440-8 (87008)
Lot Number1E31A
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeNA
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key348997
ManufacturerRICHARD-ALLAN SCIENTIFIC
Manufacturer Address225 PARSONS STRRET KALAMAZOO MI 490073538 US


Patients

Patient NumberTreatmentOutcomeDate
101. Life Threatening 2001-11-05

© 2026 FDA.report