AFFIRM VPIII AMBIENT TEMPERATURE TRANSPORT 446255

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00,06 report with the FDA on 2014-01-30 for AFFIRM VPIII AMBIENT TEMPERATURE TRANSPORT 446255 manufactured by Bd Infusion Therapy Systems Inc. S.a. De C.v..

Event Text Entries

[4166063] Customer reported that a tech was injured while running a patient sample on the affirm assay. The tech cut her finger on a glass shard inside the specimen tube. The wound required basic first aid and the tech will receive follow-up testing for (b)(6).
Patient Sequence No: 1, Text Type: D, B5


[11613755] The affirm atts collection kit is used to stabilize the nucleic acid of candida species, gardnerella vaginalis, and trichomonas vaginalis during specimen transport at ambient temperature (15-30? C) or (2-8? C) for up to 72h. The customer complaint could not be confirmed. No returns were available and no batch number was provided for investigation. An inspection of the use of the product, as directed by the insert, it appears as though the collection site took the dropper tip off when dispensing the atts solution into the sample tube. This is against the procedure and is not to be done. Bd recommends that the collection site follow the procedure as directed in the product insert in order to avoid future injury. Bd qa will continue to monitor. A review of past complaints for this product does not indicate a confirmed trend for this issue. Device was not returned to manufacturer for investigation, see manufacturer narrative below.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number0001119779-2014-00007
MDR Report Key3598848
Report Source00,06
Date Received2014-01-30
Date of Report2014-01-23
Date of Event2014-01-06
Date Mfgr Received2014-01-06
Date Added to Maude2014-01-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCHARLOTTE DANNENFELSER
Manufacturer Street7 LOVETON CIRCLE
Manufacturer CitySPARKS MD 21152
Manufacturer CountryUS
Manufacturer Postal21152
Manufacturer Phone4103164670
Manufacturer G1BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
Manufacturer StreetPERIFERICO LUIS DONALDO COLOSIO NO. 579
Manufacturer CityNOGALES 48
Manufacturer CountryMX
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAFFIRM VPIII AMBIENT TEMPERATURE TRANSPORT
Generic NameDEVICE, SPECIMEN COLLECTION
Product CodeLIO
Date Received2014-01-30
Catalog Number446255
Lot NumberUNKNOWN
OperatorMEDICAL TECHNOLOGIST
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
Manufacturer AddressPERIFERICO LUIS DONALDO COLOSIO NO. 579 NOGALES ?84048 MX ?84048


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2014-01-30

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.