AIM LED DOUBLE CEILING ALEDDC

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2013-11-26 for AIM LED DOUBLE CEILING ALEDDC manufactured by Philips Burton.

Event Text Entries

[4169596] On (b)(4) 2013, philips burton received a complaint stating that one of our double ceiling aim led light heads, had detached from the extension arm. It was reported that the head of the exam light involved, detached from the extension arm and struck a patient in the hand. No injury reported.
Patient Sequence No: 1, Text Type: D, B5


[11492685] (b)(4). (repair) documents a (b)(4) technician (philips electrical contractor) had made a visit to the site and discovered that the locking clip was not installed in the unit as referenced in the installation manual. This confirmed an installation error, made by the installers, that were either contractors or employees of (b)(6) hospital medical center. The (b)(4) contractors report ((b)(4)) to philips, stated that the light head separated from the unit because the locking clip was not installed. The device was reinstalled with the safety clip in place and the collar rotated to ensure that the clip does not come off the device. Pictures the mera technician had taken verifies that this was correctly performed. The report of this repair was submitted to philips on (b)(4) 2013.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3009542956-2013-00008
MDR Report Key3599116
Report Source05
Date Received2013-11-26
Date of Report2013-11-22
Date Mfgr Received2013-11-04
Device Manufacturer Date2013-02-25
Date Added to Maude2014-03-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactTOM MATIONG
Manufacturer Street11500 MELROSE AVE
Manufacturer CityFRANKLIN PARK IL 60131
Manufacturer CountryUS
Manufacturer Postal60131
Manufacturer Phone8472888948
Single Use3
Remedial ActionOT
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameAIM LED DOUBLE CEILING
Generic NameAIM LED DOUBLE CEILING
Product CodeFQP
Date Received2013-11-26
Model NumberALEDDC
Catalog NumberALEDDC
Lot NumberCREATE DATE: 02/25/2013
ID NumberISSUE DATE: 03/20/2013
OperatorOTHER
Device AvailabilityN
Device AgeNA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerPHILIPS BURTON
Manufacturer Address11500 MELROSE AVE FRANKLIN PARK IL 60131 US 60131


Patients

Patient NumberTreatmentOutcomeDate
10 2013-11-26

© 2026 FDA.report